_AlertIE

Fluke | 3000FC

1000V/400mA handheld digital multimeter,Fluke Conect, base for Fluke Connect modules

Fluke
Reference : FLU3000FC
Serie : 3000, Fluke-Connect
Purchase Order Reference : FLK-3000FC
Brand : Fluke
Multimètre numérique portable.
Communication sans-fil Fluke Connect.
Sert de base pour les modules Fluke Connect.
Affichage simultané jusqu'à 3 mesures maximum provenants de modules Fluke Connect, en plus des mesures standards du multimètre.
Différents modules disponibles (options): Mesure de tension, courant, température.
Portée maximum: 20 mètres. 

Gammes de mesures:
Tension AC: 1000V (multimètre ou module).
Tension DC: 1000V (multimètre).
Courant AC: 400mA (multimètre), 400A ou 2.5kA (modules).
Courant DC: 400mA (multimètre).
Mesures de température via module.
Mesures de continuité, résistance, diodes, capacité, fréquence.
Enregistrement Min/Max.
Sécurité: CAT III 1000V, CAT IV 600V.
IP54.
TTCHT
Quantity
En Stock
Quick quote for this product
Quick quote for this product
Details

Built with Fluke Connect™

See it. Save it. Share it.

All the facts. right in the field. The Fluke 3000 FC True-RMS Wireless Multimeter with the Fluke Connect™ app and ShareLive™ video call.

With its precision. reliability and ease of use. the Fluke 3000 FC Wireless Multimeter is the ideal solution for professional technicians. Independently tested for safe use in CAT IV 600V/ CAT III 1000 V environments. the Fluke 3000 FC has all the features you need to troubleshoot and repair many problems in electrical and electronic systems.

The Fluke 3000 FC gives you true-RMS voltage and current measurements. 6000-count resolution. manual and automatic ranging and provide frequency. capacitance. resistance. continuity and diode measurements. In addition. the Fluke 3000 FC provides higher 0.09% basic accuracy. a digital display showing up to three secondary measurements from remote modules. and a bright white backlight.

Work safely near electrical panels while wearing less PPE

The Fluke 3000 FC True-RMS Wireless Multimeter and Fluke Connect wireless test tools put tools. not your body. near live electrical panels. It's simple. Just de-energize the cabinet. open the panel while wearing PPE and connect the remote modules. whether they are voltage modules. current clamps. flexible current loops or thermometers. Then read the results on the 3000 FC Wireless Multimeter from a safe distance. You can display the meter measurement. plus readings from up to 3 wireless modules. while reducing the risk of arc flash by separating yourself from hazardous measurement situations.

Better still. the Fluke 3000 FC Series Wireless Multimeter can send measurement data to your smartphone. so you can save and share measurements from the field with your team anytime. from anywhere.

Don't get bogged down while chasing interrelated events

Timing is everything when trying to find the cause of a complex problem. Often you need to make a measurement in one place and see activity on a different part of the system at the same time. We've solved that problem with Fluke Connect - connect your remote modules in one location. then take your measurements up to 20 meters away with the Fluke 3000 FC multimeter. You'll be able to see the interaction between your remote test points on one single screen. in real time. You can even download the data to a PC for further analysis.

Other capabilities of Fluke Connect and the FC 3000 Multimeter

  • Add your laptop and go from logging to analysis and diagnosis.
  • Record over time using the Fluke Connect wireless modules and monitor circuit load changes for an hour. a shift or a week.
  • Use the Fluke Connect wireless USB adapter to collect logged data from remote modules by walking past a working module and downloading logged data.
  • Perform for analysis with your PC and share the results using Fluke Cloud™ storage and ShareLive™ video call. View data or graphs to get input from team members

The Fluke 3000 FC Series Wireless Multimeter with the Fluke Connect™ app has all the essentials for convenient test and measurement troubleshooting.

  • AC and DC voltage measurements to 1000V
  • AC and DC current with 0.01 mA resolution
  • Continuity. resistance. diode test. capacitance and frequency measurements
  • MIN/MAX Recording
  • CAT III 1000 V. Cat IV 600 V; IP54

Plus. the Fluke Connect wireless enabled modules measure AC. DC and AC+DC voltage. AC and DC current. plus temperature. all of which display on the Fluke 3000 FC Series Wireless Multimeter. Choose a standard clamp or flexible clamp to measure AC current. Mix and match modules to suit your unique measurement needs. Read the primary measurement on the main display and up to three modules at any one time. Modules are available separately or as part of Fluke Connect kits.

* For all specifications: accuracy is specified for one year after calibration. at operating temperatures of 18 °C to 28 °C. with relative humidity at 0 % to 90 %. Accuracy specifications take the form of ± ([ % of Reading ] + [ Number of least significant digits ]).

Detailed Specifications

AC Voltage
Range ¹ / Resolution600.0 mV / 0.1 mV
6.000 V / 0.001 V
60.00 V / 0.01 V
600.0 V / 0.1 V
1000 V / 1 V
Accuracy 2 3 445 Hz to 500 Hz1.0 % + 3
500 Hz to 1 kHz2.0% + 3
1 All AC voltage ranges are specified from 1 % of range to 100 % of range.
2 Crest factor of ≤ 3 at full scale up to 500 V. decreasing linearly to crest factor < 1.5 at 1000 V.
3 For non-sinusoidial waveforms. add ± (2 % of reading + 2 % full scale) typical. for crest factor up to 3.
4 Do not exceed 107 V-Hz.
DC Voltage. Continuity. Resistance. Diode Test and Capacitance
Function
mVRange / Resolution600.0 mV / 0.1 mV
Accuracy0.09 % + 2
VRange / Resolution6.000 V / 0.001 V
60.00 V / 0.01 V
600.0 V / 0.1 V
1000 V / 1 V
Accuracy0.09 % + 2
0.15 % + 2
)))Range / Resolution600 Ω / 1 Ω
AccuracyMeter beeps at < 25 Ω. beeper detects opens or shorts of 250 μs or longer.
Range / Resolution600.0 Ω / 0.1 Ω
6.000 kΩ / 0.001 kΩ
60.00 kΩ / 0.01 kΩ
60.00 kΩ / 0.1 kΩ
600.0 kΩ / 0.001 MΩ
50.00 MΩ / 0.01 MΩ
Accuracy0.5 % + 2
0.5 % + 1
1.5 % + 3
Diode testRange / Resolution2.000 V / 0.001 V
Accuracy1 % + 2
µFRange / Resolution1000 nF / 1 nF
10.00 μF / 0.01 μF
100.0 μF / 0.1 μF
9999 μF1 / 1 μF
Accuracy1.2 % + 2
10 % typical
1 In the 9999 μF range for measurements to 1000 μF. the measurement accuracy is 1.2 % + 2.
AC and DC Current
Function
mA AC
(45 Hz to 1 kHz)
Range 1 / Resolution60.00 mA / 0.01 mA
400.0 mA 3/ 0.1 mA
Accuracy1.5 % + 3
mA DC 2Range 1 / Resolution60.00 mA / 0.01 mA
400.0 mA 3/ 0.1 mA
Accuracy0.5 % + 3
1 All AC current ranges are specified from 5 % of range to 100 % of range.
2 Input burden voltage (typical): 400 mA input 2 mV/mA.
3 400.0 mA accuracy specified up to 600 mA overload.
Frequency
Range / Resolution99.99 Hz / 0.01 Hz
999.9 Hz / 0.1 Hz
9.999 kHz / 0.001 kHz
99.99 kHz / 0.01 kHz
Accuracy 10.1 % + 1
1 Frequency is specified up to 99.99 kHz in volts and up to 10 kHz in amps.
Input Characteristics
FunctionOverload
protection
Input impedance
(nominal)
Common mode rejection ratio (1 kΩ unbalance)Normal mode rejection
1100 V RMS> 10 MΩ < 100 pF> 120 dB at DC. 50 Hz or 60 Hz> 60 dB at 50 Hz or 60 Hz
1100 V RMS> 10 MΩ < 100 pF> 60 dB. DC to 60 Hz> 60 dB at 50 Hz or 60 Hz
1100 V RMS> 10 MΩ < 100 pF> 120 dB at DC. 50 Hz or 60 Hz> 60 dB at 50 Hz or 60 Hz
Open circuit test voltageFull scale voltageTypical short circuit current
To 6 MΩ50 MΩTypical short circuit current
1100 V RMS< 2.7 V DC< 0.7 V DC< 0.9 V DC< 350 mA
1100 V RMS< 2.7 V DC2.000 V DC< 1.1 mA
mA Function
Overload protectionFused. 44/100 A. 1000 V FAST Fuse
Overload600 mA overload for 2 minutes maximum. 10 minutes rest minimum
MIN/MAX Recording Accuracy
DC functionsThe specified accuracy of the measurement function ± 12 counts for changes > 350 mS in duration.
AC functionsThe specified accuracy of the measurement function ± 40 counts for changes > 900 mS in duration.
General Specifications
Maximum voltage between any terminal and earth ground1000 V DC or AC RMS
Ω fuse protection from A inputs0.44 A (44/100 A. 440 mA). 1000 V FAST Fuse. Fluke specified part only
Display (LCD)Update rate4/sec
Volts. amps. ohms6000 counts
Frequency10.000 counts
Capacitance1.000 counts
Battery typeThree AA Alkaline batteries. NEDA 15A IEC LR6
Battery life250 hours minimum
RF communications2.4 GHZ ISM Band
RF communication rangeOpen air. unobstructedUp to 20m
Obstructed. sheetrock wallUp to 6.5m
Obstructed. concrete wall. or steel electrical enclosureUp to 3.5m
TemperatureOperating-10 °C to 50 °C
Storage-40 °C to 60 °C
Temperature coefficient0.1 X (specified accuracy) /°C (<18 c="" or="">28 °C)
Relative humidity0 % to 90 % (0 °C to 35 °C)
0 % to 75 % (35 °C to 40 °C)
0 % to 45 % (40 °C to 50 °C)
AltitudeOperating2.000 m
Storage12.000 m
Electromagnetic compatibility
EMI. RFI. EMC. RF
EN 61326-1:2006. EN 61326-2-2:2006
ETSI EN 300 328 V1.7.1:2006. ETSI EN 300 489
V1.8.1:2008.
FCC Part 15 Subpart C Sections 15.207. 15.209. 15.249
FCCID : FCC: T68-FDMMBLE IC: 6627A-FDMMBLE
Safety complianceANSI/ISA 61010-1 / (82.02.01): 3rd edition
CAN/CSA-C22.2 No 61010-1-12: 3rd edition
UL 61010-1: 3rd edition
IEC/EN 61010-1:2010
CertificationsCSA. FCC. CE
Ingress Protection (IP) ratingIP54
Pollution Degree2
Size (H x W x L)4.75 cm x 9.3 cm x 20.7 cm (1.87 in x 3.68 in x 8.14 in)
Weight340 g (12 oz)
Note: Not compatible with Fluke CNX test tools
Product Details
41-Standard Manufacturer Warranty 3 YEARS
27- Resolution in nb points nc
33-additional measurements Resistance | Capacitance | Frequency | Temperature | Continuity test | Diode test | Min/Max
424-TRMS No
435-Data logger on PC No
436-Functions Backlight | Fuse protection
449-Direct input for current probe No
462-Current AC (Alternative) | DC (Continuous)
463-Voltage AC (Alternative) | DC (Continuous)
23-Voltage resolution 0.1mV - 1V
34-Bandwidth 45Hz - 1kHz
45-PC interface Bluetooth
24-Accuracy (%) 0.09% (DC voltage)
37-Max. DC voltage 1000 V DC
38-Max. AC voltage 1000 V AC
25-Autonomy 250 hours
39-Max. DC current 400mA DC
40-Max. AC current 400mA AC; 400A AC (module A3000); 2.5kA AC (Module i3000)
119-Safety Level CAT I | CAT II | CAT III | CAT IV
120-Safety Overvoltage CAT III 1000V / CAT IV 600V
36-Additional Characteristics Fluke Connect
132-Package includes Test leads (TL175), alligator grips (AC175), user manual
35-Dimension (HxWxD) (mm) 207x93x47.5 mm
13-Net Weight 340 g